Active Comparator Healthy Control subjects
|
|
-
Device: Pupillography
A hand-held pupillometer/electroretinogram device (RETeval, LKC) will be held in front of
the subject's eye, but will not touch the eye. The device will provide a brief, a series
of brief light stimuli and then record the pupil response and the elicited electrical
response from the retina from a surface skin patch (electrode) placed below each eye,
from the light as a measure of whether the inherent sensitivity of the eye in the retina
is normal. The investigators will repeat this in the left eye. The visible light stimulus
is safe and is given at an intensity experienced in normal daily light exposures. The
test takes about 2 minutes per eye.
Other names:
-
Device: Ocular Coherence Tomography (OCT)
The thickness of the optic nerve and macula will also be measured inside of the eye using
a special camera that forms an image of the layers of the retina without pupil dilation.
The imaging is harmless and measures the structural health of the optic nerve and retinal
layers. This test takes 5-10 minutes per eye.
-
Device: Wrist-watch sensor device
A wrist-watch sensor device (E4, Empatica) will be place on each wrist to measure skin
conductance, heart rate, skin temperature and arm movement during testing. These
wrist-watch devices are being used to monitor changes in sympathetic nerve activity to
light intensity, (the sympathetic nerves supply the blood vessels to the skin and heart).
Other names:
-
Device: Videography
The subject will sit comfortably in front of miniature combination infrared/visible light
video cameras and infrared diode light source located within 1 meter to provide video
recording of the face during testing with light and during darkness, described next.
After the 10 minutes of dark-adapting, the subject will put his/her chin on a chin rest
in front of the video cameras and a light emitting diode (LED) array give diffuse red,
blue, and white stimuli over a range of intensities. None of the stimuli are as bright as
a flash from a camera and are in the range of intensities normally experienced during
daily activities.
At the end of the test the investigators will add filters over the glasses: orange
(blue-blocking) filters and neutral density filters. Subjects will grade independently,
both the brightness and discomfort they feel from each light stimulus intensity.
-
Device: Electrophysiology
Next, electrodes will be placed above, below and to the side of the test eye to record
the electromyogram (EMG) for measuring eyelid opening and blink rate.
|
Active Comparator TBI Patients without photosensitivity
|
|
-
Device: Pupillography
A hand-held pupillometer/electroretinogram device (RETeval, LKC) will be held in front of
the subject's eye, but will not touch the eye. The device will provide a brief, a series
of brief light stimuli and then record the pupil response and the elicited electrical
response from the retina from a surface skin patch (electrode) placed below each eye,
from the light as a measure of whether the inherent sensitivity of the eye in the retina
is normal. The investigators will repeat this in the left eye. The visible light stimulus
is safe and is given at an intensity experienced in normal daily light exposures. The
test takes about 2 minutes per eye.
Other names:
-
Device: Ocular Coherence Tomography (OCT)
The thickness of the optic nerve and macula will also be measured inside of the eye using
a special camera that forms an image of the layers of the retina without pupil dilation.
The imaging is harmless and measures the structural health of the optic nerve and retinal
layers. This test takes 5-10 minutes per eye.
-
Device: Wrist-watch sensor device
A wrist-watch sensor device (E4, Empatica) will be place on each wrist to measure skin
conductance, heart rate, skin temperature and arm movement during testing. These
wrist-watch devices are being used to monitor changes in sympathetic nerve activity to
light intensity, (the sympathetic nerves supply the blood vessels to the skin and heart).
Other names:
-
Device: Videography
The subject will sit comfortably in front of miniature combination infrared/visible light
video cameras and infrared diode light source located within 1 meter to provide video
recording of the face during testing with light and during darkness, described next.
After the 10 minutes of dark-adapting, the subject will put his/her chin on a chin rest
in front of the video cameras and a light emitting diode (LED) array give diffuse red,
blue, and white stimuli over a range of intensities. None of the stimuli are as bright as
a flash from a camera and are in the range of intensities normally experienced during
daily activities.
At the end of the test the investigators will add filters over the glasses: orange
(blue-blocking) filters and neutral density filters. Subjects will grade independently,
both the brightness and discomfort they feel from each light stimulus intensity.
-
Device: Electrophysiology
Next, electrodes will be placed above, below and to the side of the test eye to record
the electromyogram (EMG) for measuring eyelid opening and blink rate.
|
Active Comparator Migraine patients without photosensitivity
|
|
-
Device: Pupillography
A hand-held pupillometer/electroretinogram device (RETeval, LKC) will be held in front of
the subject's eye, but will not touch the eye. The device will provide a brief, a series
of brief light stimuli and then record the pupil response and the elicited electrical
response from the retina from a surface skin patch (electrode) placed below each eye,
from the light as a measure of whether the inherent sensitivity of the eye in the retina
is normal. The investigators will repeat this in the left eye. The visible light stimulus
is safe and is given at an intensity experienced in normal daily light exposures. The
test takes about 2 minutes per eye.
Other names:
-
Device: Ocular Coherence Tomography (OCT)
The thickness of the optic nerve and macula will also be measured inside of the eye using
a special camera that forms an image of the layers of the retina without pupil dilation.
The imaging is harmless and measures the structural health of the optic nerve and retinal
layers. This test takes 5-10 minutes per eye.
-
Device: Wrist-watch sensor device
A wrist-watch sensor device (E4, Empatica) will be place on each wrist to measure skin
conductance, heart rate, skin temperature and arm movement during testing. These
wrist-watch devices are being used to monitor changes in sympathetic nerve activity to
light intensity, (the sympathetic nerves supply the blood vessels to the skin and heart).
Other names:
-
Device: Videography
The subject will sit comfortably in front of miniature combination infrared/visible light
video cameras and infrared diode light source located within 1 meter to provide video
recording of the face during testing with light and during darkness, described next.
After the 10 minutes of dark-adapting, the subject will put his/her chin on a chin rest
in front of the video cameras and a light emitting diode (LED) array give diffuse red,
blue, and white stimuli over a range of intensities. None of the stimuli are as bright as
a flash from a camera and are in the range of intensities normally experienced during
daily activities.
At the end of the test the investigators will add filters over the glasses: orange
(blue-blocking) filters and neutral density filters. Subjects will grade independently,
both the brightness and discomfort they feel from each light stimulus intensity.
-
Device: Electrophysiology
Next, electrodes will be placed above, below and to the side of the test eye to record
the electromyogram (EMG) for measuring eyelid opening and blink rate.
|
Active Comparator Migraine patients with photosensitivity
|
|
-
Device: Pupillography
A hand-held pupillometer/electroretinogram device (RETeval, LKC) will be held in front of
the subject's eye, but will not touch the eye. The device will provide a brief, a series
of brief light stimuli and then record the pupil response and the elicited electrical
response from the retina from a surface skin patch (electrode) placed below each eye,
from the light as a measure of whether the inherent sensitivity of the eye in the retina
is normal. The investigators will repeat this in the left eye. The visible light stimulus
is safe and is given at an intensity experienced in normal daily light exposures. The
test takes about 2 minutes per eye.
Other names:
-
Device: Ocular Coherence Tomography (OCT)
The thickness of the optic nerve and macula will also be measured inside of the eye using
a special camera that forms an image of the layers of the retina without pupil dilation.
The imaging is harmless and measures the structural health of the optic nerve and retinal
layers. This test takes 5-10 minutes per eye.
-
Device: Wrist-watch sensor device
A wrist-watch sensor device (E4, Empatica) will be place on each wrist to measure skin
conductance, heart rate, skin temperature and arm movement during testing. These
wrist-watch devices are being used to monitor changes in sympathetic nerve activity to
light intensity, (the sympathetic nerves supply the blood vessels to the skin and heart).
Other names:
-
Device: Videography
The subject will sit comfortably in front of miniature combination infrared/visible light
video cameras and infrared diode light source located within 1 meter to provide video
recording of the face during testing with light and during darkness, described next.
After the 10 minutes of dark-adapting, the subject will put his/her chin on a chin rest
in front of the video cameras and a light emitting diode (LED) array give diffuse red,
blue, and white stimuli over a range of intensities. None of the stimuli are as bright as
a flash from a camera and are in the range of intensities normally experienced during
daily activities.
At the end of the test the investigators will add filters over the glasses: orange
(blue-blocking) filters and neutral density filters. Subjects will grade independently,
both the brightness and discomfort they feel from each light stimulus intensity.
-
Device: Electrophysiology
Next, electrodes will be placed above, below and to the side of the test eye to record
the electromyogram (EMG) for measuring eyelid opening and blink rate.
|
Active Comparator TBI patients with photosensitivity
|
|
-
Device: Pupillography
A hand-held pupillometer/electroretinogram device (RETeval, LKC) will be held in front of
the subject's eye, but will not touch the eye. The device will provide a brief, a series
of brief light stimuli and then record the pupil response and the elicited electrical
response from the retina from a surface skin patch (electrode) placed below each eye,
from the light as a measure of whether the inherent sensitivity of the eye in the retina
is normal. The investigators will repeat this in the left eye. The visible light stimulus
is safe and is given at an intensity experienced in normal daily light exposures. The
test takes about 2 minutes per eye.
Other names:
-
Device: Ocular Coherence Tomography (OCT)
The thickness of the optic nerve and macula will also be measured inside of the eye using
a special camera that forms an image of the layers of the retina without pupil dilation.
The imaging is harmless and measures the structural health of the optic nerve and retinal
layers. This test takes 5-10 minutes per eye.
-
Device: Wrist-watch sensor device
A wrist-watch sensor device (E4, Empatica) will be place on each wrist to measure skin
conductance, heart rate, skin temperature and arm movement during testing. These
wrist-watch devices are being used to monitor changes in sympathetic nerve activity to
light intensity, (the sympathetic nerves supply the blood vessels to the skin and heart).
Other names:
-
Device: Videography
The subject will sit comfortably in front of miniature combination infrared/visible light
video cameras and infrared diode light source located within 1 meter to provide video
recording of the face during testing with light and during darkness, described next.
After the 10 minutes of dark-adapting, the subject will put his/her chin on a chin rest
in front of the video cameras and a light emitting diode (LED) array give diffuse red,
blue, and white stimuli over a range of intensities. None of the stimuli are as bright as
a flash from a camera and are in the range of intensities normally experienced during
daily activities.
At the end of the test the investigators will add filters over the glasses: orange
(blue-blocking) filters and neutral density filters. Subjects will grade independently,
both the brightness and discomfort they feel from each light stimulus intensity.
-
Device: Electrophysiology
Next, electrodes will be placed above, below and to the side of the test eye to record
the electromyogram (EMG) for measuring eyelid opening and blink rate.
|